Guide/Miniscope V4 Disassembly

From Miniscope
Jump to: navigation, search


Guide · 17 November 2026 20:00:00
How to take apart a Miniscope V4 to repair it, clean it, or swap parts: the tools you need, the order to work in, and which parts need the most care.

Overview

This guide covers taking apart an assembled Miniscope V4. Common reasons to open a scope are:

  • replacing a damaged coax cable or electrowetting lens (EWL)
  • removing dust or debris from the emission filter or imaging sensor
  • swapping filters or the LED to image a different fluorophore (see Converting a Miniscope V4 for a different fluorophore)
  • fixing an orientation mistake made during assembly
  • moving a working rigid-flex PCB into a new housing

Disassembly is assembly in reverse, with one difference: some joints are glued. Screwed joints come apart easily. Glued joints, where the imaging sensor PCB sits on the emission module and where the side PCBs are folded against the body, are where most damage happens.

Only go as far as you need to. Each step exposes more optics to dust and puts more wear on the screw holes. Use the table in How far to go to decide where to stop.

Exploded view of the Miniscope V4. Disassembly works from the outside of this diagram inward.
Exploded view of the Miniscope V4. Disassembly works from the outside of this diagram inward.
Video reference. There is no dedicated disassembly video yet. The assembly section of the 2021 Virtual Miniscope Workshop, Day 1 part 4 (Federico Sangiuliano) shows the tools, the exploded view, and how each module goes together, which is what you need to know to take it apart. Useful timestamps: tools (12:41), glues (13:20), exploded view (13:55), excitation module (16:43), emission module (19:00), cleaning surfaces (28:57). The per-step assembly videos are embedded in the assembly guide.
2021 Virtual Miniscope Workshop: Miniscope assembly (starts at the tools slide)

General recommendations

  • Test and write down the problem first. Record a short clip and note what is wrong (no image, dust spot, no focus change, bright square). Once the scope is open you cannot go back and check.
  • Photograph the scope from every side before you start, and again after each step. This is the easiest way to get flex routing and part orientation right on reassembly.
  • Unplug the scope from the DAQ before touching anything. Do not open a powered scope.
  • Work over lens tissue on a clean, light-colored surface, with gloves on. Small parts fall out without warning at several steps (half-ball lens, filters, EWL). A light surface makes them easy to find.
  • Have a labeled container ready for every part, and mark which side of each filter and lens faced up (see Keeping track of orientation).
  • Never pry with metal against glass, the sensor, or the flex. Pry only on Delrin or on the rigid parts of the PCB.
  • Do not use heat on the housing. Delrin softens at temperatures a heat gun reaches easily, and the press-fit lens holes will no longer hold.

Tools

Tool Needed for Notes
T2 Torx screwdriver All screwed joints Same driver as assembly. McMaster-Carr 52995A24. A worn tip strips the M1 screw heads; replace it if it slips.
Precision plastic tweezers Filters, lenses, tape Never use metal tweezers on optical surfaces. Hold parts by their edges.
Scalpel or hobby knife (#11 blade) Glue joints, coax epoxy For scoring glue lines. Cut along the joint, never toward the sensor or across the flex.
Thin plastic spudger or pry tool Lifting glued PCBs Lets you lever a PCB off without scratching. A wooden toothpick works for small gaps.
u.FL extraction tool Coax connector Hirose U.FL-LP-N-2 or similar. Pulls the connector straight up. A fingernail under the connector body also works; never pull on the cable.
Flat-tipped cotton swabs, toothpicks, wooden sticks Pushing lenses and filters out Wood and cotton will not scratch glass. The same swabs used in assembly.
Lens cleaning tissue Everything A clean work surface for parts, and for cleaning optics.
Isopropyl alcohol (IPA) Cleaning glue residue and optics Use a small amount on a swab or tissue. Keep it off the EWL and away from the sensor face.
Gloves Everything Keep skin oils off optics and the sensor.
Labeled containers (small tubes, gel-sticky boxes, or a parts tray) Storing parts One container per part. Write down the coated side of each filter.
Magnifier or stereo microscope Glue joints, flex, connector Strongly recommended for the glued steps.
Bright LED flashlight Finding dropped parts Shine it low across the bench; small lenses catch the light.
Compressed air (optional) Clearing debris Hold the can upright and spray from a distance. Do not aim at the EWL or the sensor at close range.
ESD wrist strap (optional) PCB handling Good practice when the rigid-flex PCB is exposed.
Electrowetting lens. Handle it by its edge only.

How delicate is each part?

Use this table to decide how much care each step needs. "Very delicate" parts are either expensive, hard to replace, or easily damaged beyond repair.

Care level Part Why, and what to avoid
Very delicate Imaging sensor (Python480) face Any scratch, fingerprint, or glue on the sensor shows up in every frame. Never touch it, never pry near it, and cover it as soon as it is exposed.
Flex sections of the rigid-flex PCB A sharp crease can crack traces inside the flex, and the failure may only show up later. Keep all bends smooth. The rigid-flex PCB is the hardest part of the scope to replace (see Rigid-Flex PCB).
u.FL connector Rated for only a few connect cycles. Pull straight up; do not rock or twist it, and never pull on the cable.
Electrowetting lens (EWL) A liquid lens. Do not squeeze it, press on its window, or put solvent on it. Hold it by the edge.
Dichroic mirror and emission filter Coatings scratch and corners chip. Dust on the emission filter shows in the image.
Half-ball lens 3 mm and easy to lose. It falls out as soon as the excitation module comes off.
Take care M1 screws and screw holes Self-threading screws cut their own threads in Delrin. Every removal wears the hole a little. Turn slowly, stop when snug on reassembly, and keep spares.
3 mm and 4 mm achromat lenses Press-fit. They can be pushed out, but this risks scratching them (about $90 each). Leave them in place unless you must replace them.
Excitation filter (ET470) A little dust is fine here, but chips and scratches are not.
Robust Delrin module bodies Will not break with normal handling. Do not heat them and do not lever against thin walls.
Aluminum objective sleeve Robust.
Coax cable jacket Robust, but the solder joints at each end are not.

Keeping track of orientation

Orientation mistakes are the most common reason a reassembled scope does not image. As each optic comes out:

  1. Put it on lens tissue in the same orientation it was in the scope (the side that faced the sensor facing up).
  2. Move it into its container in that orientation, and write on the container which side is up.
  3. If you lose track, re-identify the coated or curved side with the methods in Coated side and lens orientation in the assembly guide.
Part Orientation in the scope
Emission filter (ET525/50m) Coated side down, toward the dichroic
Dichroic (T495LpXR) Coated side down, toward the objective and half-ball lens
Excitation filter (ET470) Coated side toward the LED
4 mm achromat (emission module) Flat side up, toward the sensor
3 mm achromats (objective module) Flat side down, toward the brain

How far to go

Goal Steps needed
Replace the coax cable Step 1 only
Replace the excitation filter or half-ball lens, or fix a "bright small square" in the image Steps 1 to 2
Replace the electrowetting lens, or fix no focus change Steps 1 to 3
Clean dust off the emission filter or sensor Steps 1 and 4 to 5 (the excitation and objective modules can stay on)
Replace the dichroic or 4 mm tube lens Steps 1 to 6
Change filter set for a new fluorophore Steps 1 to 6, plus the LED swap in the fluorophore guide
Move the PCB to a new housing All steps (objective lenses can stay in the old housing)

Disassembly steps

Step 1: Disconnect the coax cable

The u.FL connector on the coax cable. Lift it straight up off the PCB.

Care level: very delicate.

  1. If there is a drop of epoxy securing the cable to the PCB, carefully cut it away with a scalpel. Cut on the cable side, away from the connector, so no debris falls into the connector.
  2. Slide a u.FL extraction tool (or a fingernail) under the connector body, not the cable.
  3. Lift straight up, perpendicular to the PCB. The connector snaps free with little force.
Do not pull on the cable or rock the connector side to side. This tears the connector pads off the PCB, and that is a hard repair.

Step 2: Remove the excitation module

Excitation module with the ET470 filter. The filter and the half-ball lens behind it can fall out when the module comes off.

Care level: robust module, delicate contents.

  1. Lay the scope on its side over lens tissue, with the excitation module facing up.
  2. Remove the two M1 screws with the T2 Torx driver. Turn slowly; stop if the driver slips and reseat it.
  3. Lift the excitation module off. The LED section of the rigid-flex PCB is now free to move; let it hang without bending it sharply.
  4. The half-ball lens sits in the pocket on the side of the emission module and is held in by the excitation module. It may stay in the pocket, stick to the module, or drop. Check all three places before moving on.
  5. If you need the excitation filter, push it out of its pocket from the edges with a cotton swab and note which side faced the LED.

Put the screws back loosely in the excitation module so they do not get lost.

Step 3: Separate the objective module and remove the EWL

EWL flex legs. The two gold rings sandwich the electrowetting lens between the objective and emission modules.

Care level: very delicate (EWL and flex legs).

Joining the objective and emission modules (watch it to see how they come apart)
  1. Hold the scope upright by the emission module.
  2. Loosen the three M1 screws in the objective module a little at a time, rotating between them, so the module comes off level. Do not remove one screw fully while the others are tight.
  3. Lower the objective module straight down. Do not twist; the EWL flex legs are sandwiched in the joint.
  4. The EWL sits in the top pocket of the objective module, between the two gold rings of the flex. Note which side faces up, then lift it out by its edge with plastic tweezers.
  5. Let the flex legs relax. Do not straighten them by hand.
If the objective module does not come free, check that all three screws are fully out and that the flex legs are not caught in the joint. Do not force it apart.

Step 4: Release the side PCBs (MCU/BNO and SER/POW)

The V4 rigid-flex PCB laid flat. The flex sections between the rigid boards are the parts to protect.

Care level: very delicate (flex joints).

These two rigid sections were folded down and glued to the side of the body during assembly.

  1. Under magnification, find the glue line between each board and the housing.
  2. Score along the glue line with a scalpel, cutting into the glue and not into the PCB.
  3. Slip a plastic spudger or toothpick under the edge of the board farthest from the flex, and lever gently against the Delrin.
  4. Once the board is free, let it swing out slowly. Do not unfold it flat; only open it as far as you need.
Never pull a board away by its flex. If the board does not release, score the glue again rather than pulling harder.

Glue notes:

  • UV glue (Norland 68) is fairly brittle. Scoring plus gentle levering usually breaks the joint cleanly.
  • Two-part epoxy is tougher. Score deeper and work around the edge in several passes.
  • Loctite 444 (a cyanoacrylate) softens with a little acetone applied on a swab at the glue line. Keep acetone well away from the optics and the EWL.

Step 5: Remove the imaging sensor PCB and emission filter

The emission filter sits in the pocket at the top of the emission module, directly under the imaging sensor.

Care level: very delicate. This is the riskiest step.

The Python480 sensor PCB is glued face-down onto the top of the emission module, directly over the emission filter.

  1. Hold the scope upright so the emission filter stays in its pocket when the sensor comes off.
  2. Under magnification, score around the glue line at the edge of the PCB, one side at a time.
  3. Lever at the corners of the rigid PCB with a plastic spudger, working around a little on each side until it lifts. Never insert anything under the sensor itself.
  4. Lift the PCB straight up. Cover the sensor right away: replace its protective film if you kept it, or rest it face-up in a closed, clean container.
  5. The emission filter is now exposed. Tip it out onto lens tissue, or lift it by the edges with plastic tweezers. It sits coated side down; keep it that way.
If you only need to clean dust, you may not have to remove the emission filter. Clean the filter and the sensor with lens tissue and a small amount of IPA, then test (see Reassembly) before gluing anything back.

Step 6: Remove the dichroic and 4 mm tube lens

Emission module: the T495LpXR dichroic sits in the diagonal slot, and the 4 mm achromat sits against an inner lip.

Care level: delicate.

  1. Tilt the emission module over lens tissue so the dichroic slides out of its diagonal slot. If it sticks, nudge its edge with a toothpick. Do not push on the coated face.
  2. Note the coated side (it faced down, toward the objective) and store it.
  3. Push the 4 mm achromat out with the flat end of a cotton swab, the same way it is removed when fixing an orientation mistake during assembly. It comes out with light pressure.

Step 7 (only if needed): Remove the 3 mm objective lenses

Objective module: two 3 mm achromats, press-fit flat side down.

Care level: take care. Skip this step unless you are replacing a lens or the housing.

The two 3 mm achromats are press-fit. To remove them:

  1. Place a folded lens tissue on a soft surface, and set the objective module on it, bottom down.
  2. From the top, push on the lens edge with a wooden stick or the flat end of a swab until it drops out onto the tissue.
  3. Repeat for the second lens.

Removed lenses can be reused if they are not scratched, but a lens hole that has been opened with a reamer may no longer hold a lens tightly.

Cleaning and storage

  • Remove old glue from the housing with a scalpel, and wipe the bonding surfaces with IPA before regluing. Do not scrape the sensor PCB near the sensor.
  • Clean optics with lens tissue and a small amount of IPA. Wipe in one direction, do not scrub.
  • Store each optic in its own labeled container with the orientation written on it.
  • Keep screws in their holes or in a labeled container. They are small and roll.

Reassembly

Follow the Miniscope V4 Assembly guide from the step that matches where you stopped. Some things are different the second time:

  • Screw holes are already threaded. Start each screw by turning it backward until it drops into the existing thread, then turn forward. This avoids cutting a new thread.
  • If a screw no longer bites, replace it with a new one. Stripped holes in Delrin often still hold a fresh screw.
  • Test before gluing. Power the scope and image the room through the emission filter, exactly as in assembly Step 11, to check for dust before you glue the sensor PCB back.
  • Recheck the orientation of every filter and lens. The confirmation check in assembly Step 5 (blue reflected from the top, yellow-green looking through) catches a flipped dichroic.
  • After reassembly, follow Initial testing of assembled miniscope and How to test and debug your Miniscope V4.

Replacement parts

All replacement parts are listed in Miniscope V4 part procurement. The ones most often needed after disassembly:

Part Source Part number
M1 × 3 mm self-threading screws McMaster-Carr 96817A704
T2 Torx driver McMaster-Carr 52995A24
3 mm Ø, 6 mm FL achromat (objective) Edmund Optics 45-089
4 mm Ø, 10 mm FL achromat (emission) Edmund Optics 63-691
3 mm half-ball lens Edmund Optics 47-269
Electrowetting lens Edmund Optics 34-282 (remove the metal housing and flex it ships with; see Electrowetting lens)
Filters (ET470/40x, ET525/50m, T495lpxr) Chroma Custom-diced; see Optical filters
Delrin modules and aluminum sleeve MiniscopeParts.com See Miniscope body parts
Coax cable (assembled or kit) OEPS See Coax cable
UV glue Norland NOA 68; alternatives in Screws and glue/epoxy

If the rigid-flex PCB itself is damaged, it is usually not available on its own; ask on the Miniscope Discussion Forum or contact the supplier of your scope.

See also