Difference between revisions of "Microscope Components v1"
(→Scope Body) |
(→Scope Body) |
||
Line 3: | Line 3: | ||
== Scope Body == | == Scope Body == | ||
− | The microscope body is made out of 5 machined pieces of Delrin plastic. These components are a main body, fliter cover, two part focusing | + | The microscope body is made out of 5 machined pieces of Delrin plastic. These components are a main body, fliter cover, two part focusing mechanism, and base plate. |
− | === | + | === Main Body === |
+ | This is the center piece of the microscope and holds the excitation filter, dichroic mirror, and drum lens. | ||
+ | === Filter Cover === | ||
+ | This thin cover holds the excitation filter and dichroic mirror in place in the main body. | ||
+ | |||
+ | === Focusing Mechanism === | ||
+ | Adjustment of the imaging plane is accomplished by moving the CMOS imaging sensor along the z-axis. The two part focusing mechanism consists of a fixed tube and a focusing slider. The fixed tube holds an achormatic tube lens and proves a railing for the focusing slider to move along. The CMOS imaging sensor is attached to the top of the focusing slider and can be locked in place along the fixed tube with the use of a set screw. | ||
+ | |||
+ | === Base Plate === | ||
+ | The base plate gets cemented to the animal's skull and allows easy attachment and removal of the microscope. The microscope and the base plate have 3 matching pairs of magnets which snap the microscope in the correct position in the base plate. A set screw on the base plate is then used to secure the microscope. | ||
== Optics == | == Optics == |
Revision as of 12:56, 16 February 2015
Here we will discuss the different components that go into making up our microscope. This page will give a general overview of the system as well as a detailed description of assembly.
Contents
Scope Body
The microscope body is made out of 5 machined pieces of Delrin plastic. These components are a main body, fliter cover, two part focusing mechanism, and base plate.
Main Body
This is the center piece of the microscope and holds the excitation filter, dichroic mirror, and drum lens.
Filter Cover
This thin cover holds the excitation filter and dichroic mirror in place in the main body.
Focusing Mechanism
Adjustment of the imaging plane is accomplished by moving the CMOS imaging sensor along the z-axis. The two part focusing mechanism consists of a fixed tube and a focusing slider. The fixed tube holds an achormatic tube lens and proves a railing for the focusing slider to move along. The CMOS imaging sensor is attached to the top of the focusing slider and can be locked in place along the fixed tube with the use of a set screw.
Base Plate
The base plate gets cemented to the animal's skull and allows easy attachment and removal of the microscope. The microscope and the base plate have 3 matching pairs of magnets which snap the microscope in the correct position in the base plate. A set screw on the base plate is then used to secure the microscope.