Difference between revisions of "Main Page"
Line 6: | Line 6: | ||
== Overview == | == Overview == | ||
− | Our miniature fluorescence microscope uses wide-field fluorescence imaging to record neural activity and network structure in awake, freely | + | Our miniature fluorescence microscope uses wide-field fluorescence imaging to record neural activity and network structure in awake, freely behavin mice. The functional design of our microscope is similar to a table top wide-field fluorescence microscope with the following key differences: |
+ | 1. All nonessential componets have been removed | ||
+ | 2. The table top objective was replaced with an imaging objective GRadient Index of Refraction (GRIN) lens | ||
+ | 3. The miniature fluorescence microscope does not use any infinity optics | ||
+ | 4. The excitation light source is a super bright LED rather than a lamp or laser | ||
+ | 5. The optical filters and dichroic mirror are custom cut to minimize their size and weight | ||
+ | 6. A small CMOS imaging sensor is used to record emission light instead of a large CCD, emCCD, or sCMOS camera. | ||
+ | 7. Adjustment of focal plane is done by moving imaging sensor not objective | ||
+ | |||
== Links to MINIscope subsystems == | == Links to MINIscope subsystems == |
Revision as of 12:32, 16 February 2015
UCLA MINIscope Resource
Welcome to the MINIscope Wiki page. If you do not have an account yet, please request one here.
Overview
Our miniature fluorescence microscope uses wide-field fluorescence imaging to record neural activity and network structure in awake, freely behavin mice. The functional design of our microscope is similar to a table top wide-field fluorescence microscope with the following key differences: 1. All nonessential componets have been removed 2. The table top objective was replaced with an imaging objective GRadient Index of Refraction (GRIN) lens 3. The miniature fluorescence microscope does not use any infinity optics 4. The excitation light source is a super bright LED rather than a lamp or laser 5. The optical filters and dichroic mirror are custom cut to minimize their size and weight 6. A small CMOS imaging sensor is used to record emission light instead of a large CCD, emCCD, or sCMOS camera. 7. Adjustment of focal plane is done by moving imaging sensor not objective